The Model D48 Detector Module contains a
photomultiplier, optics, filter wheel and power supply in a compact assembly (7.5 inches
wide X 6 inches high X 13 inches deep) designed for multiwavelength photon counting.
It weighs 13.5 pounds. The C&L Dye Fluorometer's innovative
filter wheel design and reasonable cost distinguish it from the competition. For a
more general discussion of this product, refer to our Product
Overview page.
Overview
- The D48 detection module incorporates a
photomultiplier operating in the photon-counting mode and a filter wheel for selection of
the detection wavelength.
- The detector module accepts light from a light
guide which interfaces to either the Model CV1 Cuvette Module
or to your fluorescence microscope using a microscope
output accessory.
- The Detector Module is controlled using the Model
PC-DAQ PC Controller. The controller interacts with the Detector Module through software
residing on your personal computer. Together, the software, PC-DAQ PC Controller and the
Model D48 Detector Module function as a complete multiwavelength data acquisition system.
- Modular design permits its use with other
illumination sources to implement multiwavelength photon counting. It is an ideal
add-on to imaging systems for versatile fluorescence quantitation.
Specifications
Modes of control
- The operation of the Model S48D is controlled
through software. Several operation modes are selectable to permit numerous
protocols from single wavelength measurements to multiwavelength measurements.
Together with the Model S48D Illumination Source, all combinations
of single and multiwavelength excitation and emission protocols are available.
- Jitter of the filter wheel when spinning is <0.5
percent (RMS) of the filter aperture time.
- Time to reposition filter wheel from one stationary
position to any other stationary position is < 60 milliseconds for an 8 position filter
wheel.
Accessible filter wheel
- The Detector Module holds either a 4 or 8 position filter wheel. The 4 position
wheel uses 1 inch diameter filters, while the 8 position wheel uses 1/2 inch diameter
filters.
- The filter wheel is easily removed from Model D48 to make rapid filter wheel changes
possible.
- Additional filter wheels can be purchased and configured for special wavelength
combinations.
Internal Shutter
- An internal shutter is provided.
- The shutter is controlled by software.
Sensitive End-On Photomultiplier
- A magnetically shielded end-on PMT is standard.
- Both bialkali and multialkali PMT versions are available.
- The output is linear to 20M Counts/sec.
- The PMT output is directly available as TTL counts for easy interface to the PC-DAQ Controller or other counting devices. Contact us for interface specifications.
Download Specification Sheet
This Detection Module may also be purchased as a
separate component for use in other systems. It can be used as a
multiwavelength detector for fluorescence video imaging systems. To interface this
detector with hardware from other manufacturers, contact
C&L Instruments. The Model D48 Detector Module, when purchased separately,
can be controlled with either the Model PC-DAQ or Model
PC-C Controller Card. These cards are similar, except that the Model PC-C Controller
does not contain data acquisition functions.
|