Click on a Topic

Model S48D Illumination Source
 

Model S48D Illumination Source

The Model S48D Illumination Source is housed in a single full width (16.73 inches wide X 6 inches high X 18.7 inches deep) rack mountable chassis. It weighs 27.5 pounds. The C&L Dye Fluorometer's innovative filter wheel design and reasonable cost distinguish it from the competition. For a more general discussion of this product, refer to our Product Overview page.

Overview

  • The Model S48D Illumination Source uses a fully digital control scheme to provide precise timing and accurate positioning of a filter wheel for wavelength selection.   It is the ideal source for multiwavelength photometry and fluorescence imaging.  
  • This illumination source requires either the PC-C, PC-DAQ, or RS-DAQ Instrument Controller to operate.  The S48D can be programmed to output up to eight wavelengths in any sequence using an eight position filter wheel. The duration of illumination at any wavelength filter position can be programmed to be continuous or to vary from 1 millisecond to several seconds.
  • This illumination source contains a light attenuation wheel for precise control of light output intensity.  Each wavelength position can be programmed to have a unique light intensity.  This is ideal for multiwavelenght and multi-dye applications.
  • The S48D contains a software-controlled shutter.
  • Adapters are available to interface the Model S48D to a variety of microscopes as well as a cuvette module.
  • Light is output using either a 3mm or 5mm core diameter quartz liquid light guide.  Fiber optic bundles are also available on request.

Specifications

Modes of control

  • The Model S48D is operated using one of three instrument controllers, the Models PC-C, PC-DAQ, or RS-DAQ Instrument Controller.  Several operation modes are selectable to permit numerous protocols from single wavelength measurements to multiwavelength measurements.  Together with the Model D48 Detector Module, all combinations of single and multiwavelength excitation and emission protocols are available.
  • Jitter of the filter wheel when spinning is <0.5 percent (RMS) of the filter aperture time.
  • Time to reposition filter wheel from one stationary position to any other stationary position is < 60 milliseconds for an 8 position filter wheel.

Accessible filter wheel

  • The S48D can hold either a 4 or 8 position filter wheel.  The 4 position wheel uses 1 inch (or 25 mm) diameter filters, while the 8 position wheel uses 1/2 inch diameter filters.
  • The filter wheel is easily removed from Model S48D to make rapid filter wheel changes possible.
  • Additional filter wheels can be purchased and configured for special wavelength combinations.

Internal Shutter

  • An internal shutter is provided.
  • The shutter is controlled by software.

150 or 300 Watt xenon arc lamp

  • Replacement lamp is a prealigned unit.
  • Capable of high power output (5.8 Watts with 300 Watt lamp).
  • Light output is from approximately 300 to 700 nm.

Control of light Intensity

  • The S48D contains a computer-controlled light attenuator wheel for precise control of output light intensity over a 50-fold range.
  • The output can be attenuated equally for all filter wavelength positions or each wavelength position can be set individually.

Download Specification Sheet